FastGene® Photometers NanoView (FG-NP02) and NanoSpec (FG-NP01) are microvolume photometers for easy and reliable biomolecule (nucleic acids and proteins) analysis.
POWERFUL • CONVENIENT • RELIABLE • COST-SAVING
Both photometers include
FastGene® NanoSpec
Spectrophotometer with a full spectrum Xenon flash lamp
The FastGene® NanoView
Photometer with LEDs measuring at three different wavelengths.
|
|
|
General specifications | ||
Photometric system | Microvolume spectrophotometer | Microvolume photometer |
Applications | Nucleic acid analysis (dsDNA, ssDNA, RNA), colorimetric protein analysis (BCA, Bradford, Lowry), cell density (OD600), kinetic tests, etc. | Nucleic acid analysis (dsDNA, ssDNA, RNA), protein analysis: absorption varies depending on the concentration at 260 nm or 280 nm (BSA, IgG), cell density (OD600) |
Footprint (W * D) | 216 mm * 290 mm | 145 mm * 190 mm |
Weight | 3.0 kg | 1.4 kg |
Optical specifications | ||
Detector | CMOS linear image sensor (2048 pixels) | Silicon photodiode |
Light source | Xenon flash lamp | LEDs |
Wavelength range | 190–850 nm (full spectrum) | 260 nm, 280 nm / 600 nm (cuvette) / 360 nm (baseline) |
Measurement time | < 8 seconds | < 10 seconds |
Absorbance accuracy | 3% (at 0.97A at 302 nm) | 3% (at 1A at 260 nm) |
Processing / Power / Software | ||
Software compatibility | Windows® 7 and 10 | – |
CPU | Octa Core ARM® CortexTM-A53 processor | |
Storage | 32 GB internal storage | 8 GB internal storage |
Connectivity | 4 x USB ports, Ethernet, and RS-232 | 2 x USB ports, USB-B, and RS-232 |
Operating system (OS) | Android | Firmware |
Display | ||
Display | 7-inch, 1280 x 800 HD color display | 4.3-inch, 480 x 272 color display |
Touchscreen | Multipoint capacitive touch | Resistive touch |
Glove compatibility | Compatible with lab gloves | Compatible with lab gloves |
Microvolume performance | ||
Minimum sample volume | 1 µL | 1 µL |
Photometric range | 0.02 – 330 A (10 mm equivalent) | 0 – 200 Abs |
Detection limit (microvolume) | 2 ng/μL (dsDNA), 0.06 mg/mL (BSA), 0.003 mg/mL (IgG) | 2 ng/μL (dsDNA) |
Maximum concentration | 16,500 ng/μL (dsDNA), 400 mg/mL (BSA) | 10,000 ng/µL (dsDNA), 298 mg/µL (BSA), 144 mg/µL (IgG) |
Cuvette performance | ||
Detection limit (cuvette) | 0.2 ng/μL (dsDNA), 0.006 mg/mL (BSA), 0.0003 mg/mL (IgG) | – |
Photometric range | 0–2 A | 0–2 A |